SRM-755 Superconducting Rock Magnetometer

Magnetómetro

The SRM-755 Superconducting Rock Magnetometer is equipment that allows the determination of the natural remanent magnetization of sediment and rock soundings, as well as discrete samples automatically, in terms of intensity and direction (magnetic moment, magnetization) and direction (spatial location vector through declination and inclination values).

Magnetómetro
Instrumental Facilities

Magnetómetro criogénico superconductor SRM-755 (2G Enterprises)
Deseñado para medir valores de magnetización débiles (nivel de ruido estimado 10-9), grazas ao seu ciclo superconductor (magnetómetro de tipo squid).

Technical characteristics:

Demagnetizing coils by alternating magnetic fields (AF) of up to 160 m

IRM (Isothermal Remanent Magnetization) magnetizer up to 1 Tesla

Anhysteretic Remanent Magnetization (ARM) Gradual Acquisition System

SQuID (Superconducting Quantum Interference Device) DC sensors: total magnetic moment dipole noise level of < 10-9 emu

and maximum measurement values ​​of 5×10-2 emu

Field attenuation 106, for transverse fields, and 1011, for axial fields

Pulse Tube Cryocooler System (completely liquid helium-free system)

Rapid cooling system (36 hours)

Access to more 4.2 cm diameter samples

 

Applications
  • Determination of natural remanent magnetization (NRM)
  • Determination of antihysteretic remanent magnetization (ARM)
  • Determination of isothermal remanent magnetization (IRM)

Very useful in the analysis of sediments and rocks in studies of palaeomagnetism and environmental magnetism, archaeomagnetism, pollen, climate change, geochronology, stratigraphy, tectonics..

Sample requirements
  • U-channels are up to 1.50 m long.
  • Discrete samples (≈10 cc)
  • One request will be made for each survey.
  • Samples will be delivered to the Core Scanning Laboratory (Floor -2, CACTI Tower Annex). Please check with the Service to confirm sample reception hours.
  • Sample collection period after analysis: 1 month

More informartion

Jorge Millos – Paula Álvarez
+34 986 81 22 38 – +34 986 81 39 68 (Core Scanning Laboratory)