Spectroquant Nova 60 Photometer

Instrumental Facilities
Spectroquant Nova 60
- Optical system: 12 filters with reference beam
- Filters (nm): 340, 410, 445, 500, 525, 550, 565, 605, 620, 665, 690, 820
- Accuracy (nm): +/- 2
- Range: -0.300 to 3.200 extinction, concentration
- Tray support: 16 mm round trays and 10, 20 and 50 mm trays
- Quality control: 3 modes (AQA)
- Data memory: 1000 data pairs
Applications
Among the 170 applications defined for drinking or wastewater, we can highlight (consult us): chlorides, fluorides, silicates, sulfides, sulfates, sulfites, cyanides, nitrates, nitrites, Total Nitroxene, phosphates, total phosphorus, ammonium, chromium, bromine, silver, anionic surfactants, phenes, etc.
Sample requirements
- Documentation to be submitted: Fully Filled Service Form
- Minimum sample quantity required: 10 mL per parameter to be determined.
- Maximum number of samples per application: 20
- Sample reception: Frozen, fixed or refrigerated (contact technical staff)
- Filtration if possible: 0.45µm
- Time frame for collecting measured samples: 1 month
- Refrigeration of analyzed samples: 1 week after delivery of results